Intrinsically safe humidity and temperature transmitter with 4-20mA output

AR003866
286,51 €
In production :  1 to 2 weeks
  • Temperature and humidity.
  • Exterior and interior.
  • Calibration certificate.
  • ATEX certification.

Volume reductions

Quantity Unit price Total price
1 286,51 € 286,51 €
10 257,86 € 2 578,60 €
25 244,97 € 6 124,25 €
50 232,07 € 11 603,50 €
Quantity

International shipping

FREE shipping in EU from 70 € HT purchase

Secure Payments

All payments are secure and approved.

Best price guarantee

Contact us to define your payment term.

+33-233-611-670

English - French - German - Spanish

TECHNICAL DATA

ATEX certified relative humidity temperature transmitter.

Outdoor and indoor use.

Integrated relative humidity, temperature sensors.

The measured values ​​are also converted into another interpretation of humidity: dew point temperature, absolute humidity, specific humidity, mixing ratio, specific enthalpy.

ATEX certification for use in potentially explosive environments. 

INCLUDED IN DELIVERY

  • T3110Ex
  • Manual
  • Traceable calibration certificate
  • Free Tsensor configuration software 


MEASURED VALUES

- Temperature
- Relative humidity
- Calculated values ​​(dew point, etc.)

Relative humidity range: 0 to 100%

Accuracy of the relative humidity measurement: ± 2,5% relative humidity from 5 to 95% at 23 ° C

Accuracy of temperature output: ± 0.4 ° C

Available temperature units: degrees Celsius, Fahrenheit

Dew point output accuracy and range - for details see graphs: ± 1,5 ° C at ambient temperature T <25 ° C and relative humidity> 30%
range -60 to +80 ° C

Dew point output accuracy and range - for details see graphs Accuracy and dew point output - for details see graphs: ± 1,5 g / m3 at ambient temperature T <25 ° C
range from 0 to 400 g / m3

Accuracy and specific humidity output range: ± 2 g / kg at room temperature T <35 ° C
range 0 to 550 g / kg

Accuracy and output range of mixing ratio: ± 2g / kg at ambient temperature T <35 ° C
range from 0 to 995 g / kg

Precision and range of specific enthalpies: ± 3 kJ / kg at ambient temperature T <25 ° C
range: 0 to 995 kJ / kg

Temperature operating range: -30 to + 60 ° C

Operating temperature range (environment of the case): up to + 60 ° C

LCD display operating temperature range: readable at an operating temperature of + 70 ° C, it is recommended to turn off the LCD screen above 70 ° C

Humidity sensor temperature compensation range: 4-20mA, galvanically isolated temperature range

Current outputs - two-wire connection: 4-20mA, galvanically isolated

Configuration of outputs and output range: user adjustable from PC

Filtering capacity of the sensor cover: 0.025mm - filter with stainless steel mesh

Housing protection with electronics: Electronics IP65, sensors IP40

Supply : 9-30Vdc

dimensions: 88,5 x 170 x 39,5 mm (W x H x D), rod length 75 mm

Weight: about 150g

Warranty: 3 years 

Information

Measured
Humidity
Communication
4 ... 20mA
Related products (2 other products in the same category)