2-wire
Both leads and contacts add to the element resistance. Assess that contribution against the required accuracy.
GUILCOR / TEMPERATURE INTELLIGENCE
Match a 120 Ω nickel sensor to the correct measuring input. Check its reference curve and use a two-point comparison to review measured resistance before commissioning.
Send the existing sensor and controller references when specifying a replacement.

Drag to rotate. Scroll or pinch to zoom.
02 / MATCH THE COMPLETE SPECIFICATION
Nominal resistance and the curve both need to match.
Ni120 denotes a nickel RTD with 120 Ω nominal resistance at 0 °C. A common instrument characteristic is the Edison-type 0.00672 curve, published by Minco as NA. It is not obtained by simply multiplying the DIN Ni100 curve by 1.2.
Nickel resistance is nonlinear. This page uses the published NA resistance table and linear interpolation between adjacent 1 °C entries. An instrument must support the same characteristic; a generic nickel setting or a DIN Ni100 setting does not establish compatibility.
Material, nominal resistance, reference temperature and the exact curve.
Controller part number, supported input, terminal arrangement and measuring current.
Operating and peak temperatures, accuracy requirement and response conditions.
Probe dimensions, insertion, mounting, cable, connector, sealing and environment.
03 / THE SPECIFIED CHARACTERISTIC
Values follow the cited table, with linear interpolation between adjacent 1 °C points. This is a nominal reference, not a calibration or a product temperature rating.
The reference is read from the published resistance table. Between two entries, R = R₁ + (T − T₁)(R₂ − R₁)/(T₂ − T₁). The inverse uses the same segment.
Displayed domain: −60 to 180 °C. Account for leads and contacts before interpreting a resistance reading.
For this named reference only. Match your sensor and input before using the values.
NI120 / PRACTICAL TOOL
Enter two known, stable temperatures and the element resistance measured at each. Compare the differences with an allowance taken from your test procedure or datasheet. Example values are provided to show the workflow.
This is a nominal resistance comparison, not a calibration certificate or an open/short diagnosis. Use a suitable resistance instrument, account for leads and contacts, and allow thermal equilibrium. The temperatures must be different and within the displayed reference domain.
04 / CONNECTIONS
Both leads and contacts add to the element resistance. Assess that contribution against the required accuracy.
A supported compensation scheme reduces lead effects when its assumptions, including lead matching, are met.
Separate force and sense conductors reduce the effect of lead voltage drops. Use the terminal assignment specified by the instrument.
05 / THE COMPLETE PROBE
Request the actual Ni120 tolerance or calibrated resistance data. A generic platinum Class A/B table is not a Ni120 accuracy specification. The two-point tool uses an allowance you supply and cannot establish performance between or beyond those points.
Electrical excitation dissipates I²R in the element. Temperature rise depends on the actual dissipation conditions. Use the selected product’s measuring-current limit and self-heating data.
Specify sheath, seals, insulation, cable and connector for moisture, pressure, chemicals and vibration. Material name alone does not define a finished probe’s rating.
Compare response times only with their medium, flow, mounting and response criterion. Request drift or stability data at the intended exposure conditions.
REFERENCE / CHECK THE NUMBERS
At 100 °C the named reference is 200.640 Ω. Reversing that same reference gives 100 °C before rounding. Product tolerance, reference uncertainty and wiring affect real measurements.
Temperature = row value + column offset. The table interval is a calculation domain, not a GUILCOR probe rating.
| Temperature °C | +0 | +1 | +2 | +3 | +4 | +5 | +6 | +7 | +8 | +9 |
|---|---|---|---|---|---|---|---|---|---|---|
| −60 | 79.620 | 80.273 | 80.926 | 81.580 | 82.234 | 82.888 | 83.542 | 84.197 | 84.853 | 85.508 |
| −50 | 86.165 | 86.821 | 87.479 | 88.136 | 88.794 | 89.453 | 90.112 | 90.772 | 91.433 | 92.094 |
| −40 | 92.755 | 93.418 | 94.080 | 94.744 | 95.408 | 96.073 | 96.739 | 97.406 | 98.073 | 98.741 |
| −30 | 99.410 | 100.080 | 100.750 | 101.422 | 102.094 | 102.768 | 103.442 | 104.117 | 104.794 | 105.471 |
| −20 | 106.150 | 106.829 | 107.510 | 108.192 | 108.876 | 109.560 | 110.246 | 110.933 | 111.622 | 112.312 |
| −10 | 113.003 | 113.696 | 114.390 | 115.086 | 115.783 | 116.482 | 117.182 | 117.884 | 118.588 | 119.293 |
| 0 | 120.000 | 120.709 | 121.419 | 122.132 | 122.846 | 123.562 | 124.279 | 124.999 | 125.720 | 126.443 |
| 10 | 127.168 | 127.895 | 128.623 | 129.353 | 130.086 | 130.820 | 131.556 | 132.294 | 133.033 | 133.775 |
| 20 | 134.518 | 135.264 | 136.011 | 136.760 | 137.512 | 138.265 | 139.020 | 139.777 | 140.536 | 141.297 |
| 30 | 142.060 | 142.825 | 143.592 | 144.361 | 145.132 | 145.905 | 146.680 | 147.457 | 148.236 | 149.017 |
| 40 | 149.800 | 150.586 | 151.373 | 152.162 | 152.954 | 153.747 | 154.543 | 155.340 | 156.140 | 156.942 |
| 50 | 157.745 | 158.551 | 159.359 | 160.170 | 160.982 | 161.796 | 162.613 | 163.431 | 164.252 | 165.075 |
| 60 | 165.900 | 166.727 | 167.557 | 168.388 | 169.222 | 170.057 | 170.895 | 171.735 | 172.577 | 173.422 |
| 70 | 174.268 | 175.117 | 175.967 | 176.820 | 177.675 | 178.532 | 179.391 | 180.253 | 181.116 | 181.981 |
| 80 | 182.849 | 183.719 | 184.590 | 185.464 | 186.340 | 187.218 | 188.098 | 188.981 | 189.865 | 190.751 |
| 90 | 191.639 | 192.530 | 193.422 | 194.317 | 195.214 | 196.113 | 197.014 | 197.917 | 198.823 | 199.730 |
| 100 | 200.640 | 201.551 | 202.465 | 203.381 | 204.299 | 205.220 | 206.142 | 207.067 | 207.994 | 208.923 |
| 110 | 209.854 | 210.788 | 211.723 | 212.661 | 213.601 | 214.544 | 215.488 | 216.435 | 217.384 | 218.335 |
| 120 | 219.290 | 220.246 | 221.204 | 222.164 | 223.127 | 224.092 | 225.059 | 226.028 | 227.000 | 227.974 |
| 130 | 228.950 | 229.928 | 230.909 | 231.892 | 232.877 | 233.865 | 234.855 | 235.847 | 236.841 | 237.838 |
| 140 | 238.836 | 239.837 | 240.841 | 241.847 | 242.855 | 243.865 | 244.877 | 245.892 | 246.909 | 247.928 |
| 150 | 248.950 | 249.974 | 251.000 | 252.029 | 253.059 | 254.093 | 255.128 | 256.166 | 257.207 | 258.250 |
| 160 | 259.295 | 260.343 | 261.394 | 262.447 | 263.503 | 264.561 | 265.622 | 266.686 | 267.753 | 268.822 |
| 170 | 269.894 | 270.968 | 272.046 | 273.126 | 274.209 | 275.296 | 276.385 | 277.476 | 278.571 | 279.669 |
| 180 | 280.770 | — | — | — | — | — | — | — | — | — |
Values follow the cited table, with linear interpolation between adjacent 1 °C points. This is a nominal reference, not a calibration or a product temperature rating. Source document ↗
ELECTRONICS / FROM RESISTANCE TO A READING
At an illustrative 0.3 mA, the reference produces 36.00 mV at 0 °C and 60.19 mV at 100 °C. Match gain and input range to the full interval, and confirm that the current meets the sensor datasheet.
A passive RTD does not provide 0–10 V or 4–20 mA by itself. Specify a compatible RTD input or transmitter. Check gain, reference, input range, common-mode limits and calibration over the intended interval.
SELECTION / YOUR APPLICATION
The equipment specifies Ni120 with a documented nickel curve. Confirm the actual reference characteristic before comparing another nominal value.
Review the specification →Compare the nominal resistance, material, reference curve and supported wiring. A shared RTD label does not make two references compatible with the same input.
Explore Cu10 →Keep the existing sensor and instrument references. Check the specified curve, tolerance interval, wire count and complete probe dimensions before choosing a replacement.
Prepare your enquiry →These links support a specification review. They do not establish interchangeable parts or availability of a complete assembly.
Ni120 suits measurement systems designed for its 120 Ω nominal value and specified nickel characteristic. It provides a defined replacement target when the original sensor and input documentation are available.
Higher nominal resistance does not automatically improve cable stability or eliminate amplification and noise requirements. Check the element curve, lead compensation, excitation and operating interval as a complete system.
Continue comparing the family
07 / APPLICATIONS
These application contexts and probe formats require confirmation for the selected sensor reference.

Keep the installed controller reference and its nickel characteristic with the replacement specification.

Check known temperature points using suitable resistance measurement and traceable references.

Define packaging, installation, cable and environmental limits for the application.
08 / FORM FACTOR LIBRARY
The Ni120 can take many forms. These are GUILCOR probe families and a design service, not a promise that every format has a Ni120 option. Confirm the construction for the selected reference.
10 / PRACTICAL ANSWERS
Not for the reference curves used here: Ni120 NA uses an average coefficient near 0.00672, while DIN Ni100 uses about 0.00618. Scaling only nominal resistance gives the wrong relationship.
No. They provide limited checks at the stated conditions. Calibration also requires appropriate references, uncertainty evaluation and a procedure covering the intended use.
No. Confirm the element, class, sheath, seals, cable and connector limits. A reference table’s domain is not the assembled probe rating.
Not automatically. Sensor tolerance, instrument performance, self-heating, wiring and thermal contact all contribute. Select the complete measurement chain.
Send the existing probe and controller references, the resistance curve, operating interval, accuracy requirement, wiring, dimensions, mounting, environment and quantity.
GUILCOR / YOUR ENQUIRY
Send the electrical reference and the installed probe requirements together. Include the instrument configuration so the sensing element and readout can be reviewed as one measurement chain.
Send the component marking, equipment reference and any photographs or documents you have. These provide a starting point for identifying what must be confirmed.
The outline helps prepare your enquiry. Copying it does not send a request.
FOUR DETAILS THAT MOVE THE DISCUSSION FORWARD
Sensor material, nominal resistance and reference temperature, curve, tolerance class and interval, instrument model and 2-, 3- or 4-wire input.
Operating and peak temperatures, required system accuracy, response target, measured medium and calibration points.
Sheath dimensions, insertion and mounting, cable length and insulation, wire count, termination and environmental exposure.
Sample and production quantities, target delivery, calibration or documentation needs, and the checks required before approval.
We use cookies to provide you a better user experience on this website. Cookie Policy