Skip to Content
Temperature sensing. From element to assembly.+33 (0)2 33 61 16 70

GUILCOR / TEMPERATURE INTELLIGENCE

Ni120
temperature sensors

Match a 120 Ω nickel sensor to the correct measuring input. Check its reference curve and use a two-point comparison to review measured resistance before commissioning.

Send the existing sensor and controller references when specifying a replacement.

Interactive 3D model of the Ni120 temperature sensor

Drag to rotate. Scroll or pinch to zoom.

3D illustration of Ni120. Confirm dimensions and connections in the documentation for your selected reference.
NickelSensing material
120 Ω at 0 °CNickel NA · α ≈ 0.00672 °C⁻¹
2 / 3 / 4 wiresConfirm the instrument’s supported connection

02 / MATCH THE COMPLETE SPECIFICATION

What is a Ni120 sensor?

Nominal resistance and the curve both need to match.

Ni120 denotes a nickel RTD with 120 Ω nominal resistance at 0 °C. A common instrument characteristic is the Edison-type 0.00672 curve, published by Minco as NA. It is not obtained by simply multiplying the DIN Ni100 curve by 1.2.

Nickel resistance is nonlinear. This page uses the published NA resistance table and linear interpolation between adjacent 1 °C entries. An instrument must support the same characteristic; a generic nickel setting or a DIN Ni100 setting does not establish compatibility.

01

Electrical reference

Material, nominal resistance, reference temperature and the exact curve.

02

Instrument & wiring

Controller part number, supported input, terminal arrangement and measuring current.

03

Measurement point

Operating and peak temperatures, accuracy requirement and response conditions.

04

Mechanical assembly

Probe dimensions, insertion, mounting, cable, connector, sealing and environment.

03 / THE SPECIFIED CHARACTERISTIC

Temperature and resistance. One reference tool.

Minco NA — 120 Ω nickel reference table

Values follow the cited table, with linear interpolation between adjacent 1 °C points. This is a nominal reference, not a calibration or a product temperature rating.

The reference is read from the published resistance table. Between two entries, R = R₁ + (T − T₁)(R₂ − R₁)/(T₂ − T₁). The inverse uses the same segment.

Read the reference source ↗

Displayed domain: −60 to 180 °C. Account for leads and contacts before interpreting a resistance reading.

Reference resistance200.640 Ω

For this named reference only. Match your sensor and input before using the values.

NI120 / PRACTICAL TOOL

Check two reference points before commissioning.

Enter two known, stable temperatures and the element resistance measured at each. Compare the differences with an allowance taken from your test procedure or datasheet. Example values are provided to show the workflow.

This is a nominal resistance comparison, not a calibration certificate or an open/short diagnosis. Use a suitable resistance instrument, account for leads and contacts, and allow thermal equilibrium. The temperatures must be different and within the displayed reference domain.

Reference point 1
Reference point 2
Enable JavaScript to explore

04 / CONNECTIONS

Match the wiring to the measuring input.

Ⅱ

2-wire

Both leads and contacts add to the element resistance. Assess that contribution against the required accuracy.

Ⅲ

3-wire

A supported compensation scheme reduces lead effects when its assumptions, including lead matching, are met.

Ⅳ

4-wire

Separate force and sense conductors reduce the effect of lead voltage drops. Use the terminal assignment specified by the instrument.

05 / THE COMPLETE PROBE

Specify accuracy, response and protection together.

Curve & tolerance

Request the actual Ni120 tolerance or calibrated resistance data. A generic platinum Class A/B table is not a Ni120 accuracy specification. The two-point tool uses an allowance you supply and cannot establish performance between or beyond those points.

Self-heating

Electrical excitation dissipates I²R in the element. Temperature rise depends on the actual dissipation conditions. Use the selected product’s measuring-current limit and self-heating data.

Construction & environment

Specify sheath, seals, insulation, cable and connector for moisture, pressure, chemicals and vibration. Material name alone does not define a finished probe’s rating.

Response & stability

Compare response times only with their medium, flow, mounting and response criterion. Request drift or stability data at the intended exposure conditions.

REFERENCE / CHECK THE NUMBERS

Ni120 reference values and a worked example.

At 100 °C the named reference is 200.640 Ω. Reversing that same reference gives 100 °C before rounding. Product tolerance, reference uncertainty and wiring affect real measurements.

Full −60 to 180 °C reference table, in 1 °C steps

Temperature = row value + column offset. The table interval is a calculation domain, not a GUILCOR probe rating.

Minco NA — 120 Ω nickel reference table — resistance in ohms
Temperature °C+0+1+2+3+4+5+6+7+8+9
−6079.62080.27380.92681.58082.23482.88883.54284.19784.85385.508
−5086.16586.82187.47988.13688.79489.45390.11290.77291.43392.094
−4092.75593.41894.08094.74495.40896.07396.73997.40698.07398.741
−3099.410100.080100.750101.422102.094102.768103.442104.117104.794105.471
−20106.150106.829107.510108.192108.876109.560110.246110.933111.622112.312
−10113.003113.696114.390115.086115.783116.482117.182117.884118.588119.293
0120.000120.709121.419122.132122.846123.562124.279124.999125.720126.443
10127.168127.895128.623129.353130.086130.820131.556132.294133.033133.775
20134.518135.264136.011136.760137.512138.265139.020139.777140.536141.297
30142.060142.825143.592144.361145.132145.905146.680147.457148.236149.017
40149.800150.586151.373152.162152.954153.747154.543155.340156.140156.942
50157.745158.551159.359160.170160.982161.796162.613163.431164.252165.075
60165.900166.727167.557168.388169.222170.057170.895171.735172.577173.422
70174.268175.117175.967176.820177.675178.532179.391180.253181.116181.981
80182.849183.719184.590185.464186.340187.218188.098188.981189.865190.751
90191.639192.530193.422194.317195.214196.113197.014197.917198.823199.730
100200.640201.551202.465203.381204.299205.220206.142207.067207.994208.923
110209.854210.788211.723212.661213.601214.544215.488216.435217.384218.335
120219.290220.246221.204222.164223.127224.092225.059226.028227.000227.974
130228.950229.928230.909231.892232.877233.865234.855235.847236.841237.838
140238.836239.837240.841241.847242.855243.865244.877245.892246.909247.928
150248.950249.974251.000252.029253.059254.093255.128256.166257.207258.250
160259.295260.343261.394262.447263.503264.561265.622266.686267.753268.822
170269.894270.968272.046273.126274.209275.296276.385277.476278.571279.669
180280.770—————————

Values follow the cited table, with linear interpolation between adjacent 1 °C points. This is a nominal reference, not a calibration or a product temperature rating. Source document ↗

ELECTRONICS / FROM RESISTANCE TO A READING

Build the complete measurement channel.

At an illustrative 0.3 mA, the reference produces 36.00 mV at 0 °C and 60.19 mV at 100 °C. Match gain and input range to the full interval, and confirm that the current meets the sensor datasheet.

Excitation→Ni120 & leads→Conditioning & ADC→Correct curve

A passive RTD does not provide 0–10 V or 4–20 mA by itself. Specify a compatible RTD input or transmitter. Check gain, reference, input range, common-mode limits and calibration over the intended interval.

SELECTION / YOUR APPLICATION

Is Ni120 the right starting point?

COMPARE BEFORE CHANGING

Review Cu10 alongside Ni120.

Compare the nominal resistance, material, reference curve and supported wiring. A shared RTD label does not make two references compatible with the same input.

Explore Cu10 →
REPLACING AN EXISTING PART

Keep the complete specification together.

Keep the existing sensor and instrument references. Check the specified curve, tolerance interval, wire count and complete probe dimensions before choosing a replacement.

Prepare your enquiry →

These links support a specification review. They do not establish interchangeable parts or availability of a complete assembly.

More selection details

Ni120 suits measurement systems designed for its 120 Ω nominal value and specified nickel characteristic. It provides a defined replacement target when the original sensor and input documentation are available.

Higher nominal resistance does not automatically improve cable stability or eliminate amplification and noise requirements. Check the element curve, lead compensation, excitation and operating interval as a complete system.

07 / APPLICATIONS

Match the assembly to the measurement point.

These application contexts and probe formats require confirmation for the selected sensor reference.

GUILCOR temperature probe format illustration

Existing nickel systems

Keep the installed controller reference and its nickel characteristic with the replacement specification.

GUILCOR temperature probe format illustration

Test benches

Check known temperature points using suitable resistance measurement and traceable references.

GUILCOR temperature probe format illustration

Thermal automation

Define packaging, installation, cable and environmental limits for the application.

08 / FORM FACTOR LIBRARY

Choose your Ni120 probe format

The Ni120 can take many forms. These are GUILCOR probe families and a design service, not a promise that every format has a Ni120 option. Confirm the construction for the selected reference.

10 / PRACTICAL ANSWERS

Practical questions about Ni120.

Is Ni120 just a scaled Ni100?

Not for the reference curves used here: Ni120 NA uses an average coefficient near 0.00672, while DIN Ni100 uses about 0.00618. Scaling only nominal resistance gives the wrong relationship.

Do two matching points prove the sensor is calibrated?

No. They provide limited checks at the stated conditions. Calibration also requires appropriate references, uncertainty evaluation and a procedure covering the intended use.

Does every Ni120 have the same temperature range?

No. Confirm the element, class, sheath, seals, cable and connector limits. A reference table’s domain is not the assembled probe rating.

Does a higher signal mean better accuracy?

Not automatically. Sensor tolerance, instrument performance, self-heating, wiring and thermal contact all contribute. Select the complete measurement chain.

What information helps with a replacement?

Send the existing probe and controller references, the resistance curve, operating interval, accuracy requirement, wiring, dimensions, mounting, environment and quantity.

GUILCOR / YOUR ENQUIRY

Define your Ni120 enquiry.

Send the electrical reference and the installed probe requirements together. Include the instrument configuration so the sensing element and readout can be reviewed as one measurement chain.

Missing a reference or datasheet?

Send the component marking, equipment reference and any photographs or documents you have. These provide a starting point for identifying what must be confirmed.

Discuss your enquiry ↗

The outline helps prepare your enquiry. Copying it does not send a request.

FOUR DETAILS THAT MOVE THE DISCUSSION FORWARD

01

Sensor and instrument

Sensor material, nominal resistance and reference temperature, curve, tolerance class and interval, instrument model and 2-, 3- or 4-wire input.

02

Measurement conditions

Operating and peak temperatures, required system accuracy, response target, measured medium and calibration points.

03

Assembly and connection

Sheath dimensions, insertion and mounting, cable length and insulation, wire count, termination and environmental exposure.

04

Project and validation

Sample and production quantities, target delivery, calibration or documentation needs, and the checks required before approval.