Temperature probe with metal connection head

Temperature sensor - Metal connection head *Step required

In production :  2 to 3 weeks
Description and application

These resistance temperature sensors are designed to measure the contact temperature of liquid or gaseous substances. The combination of temperature sensor and central support is suitable for temperature measurements in air conditioning ducts. The combination of temperature sensor and thermowell is suitable for temperature measurements in tubes. The sensor variant with welded thread is ideal for direct measurement of fluids in ducts. The standard operating temperature range is between -30 ° C and 200 ° C. By using a sensor with a longer rod, the upper allowable temperature limit can be extended up to 250 ° C. The temperature sensors can be used for any control system compatible with the output signals of the sensing elements or the output signals shown in the table of sensing element types.

The temperature sensors are designed to operate in a chemically non-aggressive environment.

Unit price
39,00 €
Final price
39,00 €
Your configuration:

International shipping

FREE shipping in Europe from 100 € of purchase

Secure Payments

All payments are secure and approved.

Best price guarantee

Contact us to define your payment term.

+33-233-611-670

English - French - German - Spanish

ACCESSORIES
  • central metal support K120
  • stainless steel thermowell JS 130
  • screw with collar or cutting rings - if different immersion lengths of the temperature sensor are set

DECLARATION, CERTIFICATION, CALIBRATION

  • The manufacturer provides an EU declaration of conformity.
  • Calibration - The final metrological control - comparison with standards or working instruments - is carried out for all products. The continuity of standards and work measurement instruments is ensured within the meaning of article 5 of Law No. 505/1990 on metrology. The manufacturer offers the possibility of supplying the calibrated sensors.
  • Features
  • Cable schematics
  • Dimensional draft
  • Download
Sensor type NK 120 NK 121 NK 122 320 NK NK 321
Sensitive element type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range -30 to 200 ° C (ambient temperature of the connection head -30 to 100 ° C)
Maximum continuous current 1 mA 1 mA 1 mA 0.3 mA 0.3 mA

Sensor type NK 123 PTK 120 PTK 220 PTK 320 HK 120
Sensitive element type T1 = Ni 2226 Pt 100/3850 Pt 500/3850 Pt 1000/3850 20 kΩ NTC thermistor
Measuring range -30 to 150 ° C - 50 to 200 ° C (ambient temperature of the connection head -30 to 100 ° C) -30 to 150 ° C
Maximum continuous current 0.7 mA 3 mA 1.5 mA 1 mA 10 mW *)


*) maximum power consumption

Sensor type NK 520 Notes
Sensitive element type Pt 1000/3850
Output signal 4 to 20 mA
Measuring ranges -50 to 50 ° C 
-30 to 60 ° C 
0 to 35 ° C 
0 to 100 ° C 
0 to 150 ° C 
0 to 200 ° C 
0 to 250 ° C
ambient temperature around the connection head -30 to 70 ° C
Measurement error <0.6% of range not less than 0.5 ° C
Power supply (U) 11 to 30 V DC recommended value 24 V DC
Load resistance 150 Ω for 12 VDC power supply 
700 Ω for 24 VDC power supply
Output signal 
breaking the sensitive element
> 24 mA
Output signal 
- short circuit of the detection element
<3.5 mA

OTHER PARAMETERS
Precision class Elements sensitive to Ni: class B, Δt = ± (0,4 + 0,007t), for t ≥ 0; Δt = ± (0,4 + 0,028 | t |), for t ≤ 0 in ° C; 
Elements sensitive to pt: class B according to EN 60751, Δt = ± (0,3 + 0,005 | t |) in ° C 
NTC 20 kΩ: ± 1 ° C for the range from 0 to 70 ° C
Sensor connection according to the wiring diagram
Standard length of rod L1 70, 120, 180, 240, 300, 360, 420 mm
Response time τ 0,5 <9 s (in white water at 0,4 ms -1 )
Recommended wire section - 
sensors with grommet
0,35 to 1,5 mm 2
Isolation resistance > 200 MOhm at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Ingress protection IP 54 according to EN 60529
Material of the stem DIN 1.4301 stainless steel
Connection head type LIMATHERM MA
Connection head material aluminum alloy
operating conditions ambient temperature: 
-30 to 100 ° C; -30 to 70 ° C with converter 
relative humidity: max. 100% (at room temperature 25 ° C) 
atmospheric pressure: 70 to 107 kPa
Weight about 0,15 kg

Information

Measured
Temperature
Product Type
Head / terminal probes
Communication
4 ... 20mA
Related products (8 other products in the same category)

Probe with standard connection head

Description and application These resistance sensors are intended for contact measurement of temperatures up to 4 ...
Price 50,89 €
2 to 3 weeks